    Displaying RGT/SMP Error on the CA610 Fluoride analyzer

    產品分類:CA610在線氟化物分析儀
    更新日期:2018-11-09
    ● 問題咨詢

    Displaying RGT/SMP Error on the CA610 Fluoride analyzer

    ● 關鍵詞摘要

    CA610,在線氟化物

    ● 解答

    RGT/SMP Error is on software Ver 1.4 and can be cleared on the CA610 Fluoride analyzer.  It may be caused by bubbles in the sample stream, or that cling inside the measurement chamber.  The message doesn't happen as often with the latest Ver 1.5 software.  If this is a reoccurring problem, and the instrument has Ver 1.4, the entire circuit board has to be replaced in order for it to update to the newest version.

    這個解答對您是否有幫助?